Table of Contents

Quick Guide

High Level View

Automatic mechanism

The following process is carried out every day:

Automatic processing

Automatic analysis works in stages to detect molecules.

Between each stage, alerts are raised based on the data calculated.

Alert system

The alerts raised by the automatic processing enable the molecule to be classified for a given sample.

Main functionalities

Login view

Authentication form

Once authenticated, redirection to the Backlog view

Direct access to the Backlog view.

Direct access to the current Batch view (only from Analysis / Compound view).

List of analyses in the current batch:

List of Compounds (molecules) in the current Batch:

:!: Navigation is not displayed from the Backlog.

Backlog view

Batch view

An overview of the entire batch and the progress of reprocessing.

Analysis view

Synthetic view of an analysis within a series.

Compound view

Synthetic view of a molecule within a series. It enables a comparison to be made between all the analyses for a specific molecule.

Validation

Validation process

:!: Quick validation is possible using multi selection

Progression

Progression of validation

Peak reintegration

Ability to integrate a peak or re-integrate an existing peak into a chromatogram

Work process

Standard work process can be the following:

  1. Select the series you want to work on
  2. Check the series
    • Checking calibrations
    • Checking blanks
    • Checking organic matrices
    • Verification of bio-doped matrices
  3. Choose a sample
    • look at the detected first, then suspected by sorting on the area, for example
  4. Select the customer sample to be checked
    • Verification and validation of detected molecules
    • Verification and validation of suspected molecules
    • Optional verification and validation of excluded molecules
  5. Update the LIMS via the Batch view